Platform – Flexia Cobalt Chromium Coronary Stent System (CE Marked)

Flexia coronary stent is next generation uniform sinusoidal strut design offering uniform drug delivery throughout stent length. Flexia offers excellent flexibility & exceptional deliverability.

Unique Features

  • Excellent Flexibility and radial strength
  • Trackability as never before
  • Everolimus drug coated
  • Reduces artery Injury
  • Next generation strut design offering excellent performance
  • Immensely thin struts to minimize the focal injury of vessel wall
  • Awesome surface finish
Unique Features - Eternia Drug Eluting Stent - Nano Therapeutics Pvt. Ltd. - Heart Stent Manufacturing Company Surat, India
Engineered to Deliver Safety and Efficacy
Engineered to Deliver Safety and Efficacy - Eternia Drug Eluting Stent - Nano Therapeutics Pvt. Ltd. - Heart Stent Manufacturing Company Surat, India
Drug Release Kinetics

Eternia has proven drug release kinetics. Initial burst release of Everolimus followed by sustained release up to 40 days.

Strut Thickness Matters

Eternia is the thinnest DES available in the market. Thin strut & polymer coating reduces injury to vessel wall and aid faster re-endotheliazation thereby minimizing the risk of Thrombosis.

Biodegradable Polymer

The Polymer completely degrades by Hydrolysis & enzymatic degradation which is eventually excreted from the body in form of CO2 and H2O

Technical Specification
Available Stent Length (mm) 8,  13, 16, 19, 24, 29, 32, 37, 40
Available Stent Diameter (mm) 2.25, 2.50, 2.75, 3.00, 3.50, 4.00, 4.50
Design Uniform sinusoidal cell design
Stent Material L605 Cobalt Chromium alloy, Stent Platform:  Flexia
Recoil <4%
Drug Component Everolimus Drug and Biodegradable Polymer
Drug Dose 1.20 µg/mm2
Coating Thickness 1.5 µm
Stent Strut thickness 65 µm
Balloon Delivery System Rapid Exchange
Hydrophilic Coating On the distal and proximal shaft of balloon delivery system.
Delivery System Usable Length 140 cm
Balloon Inflation Pressure Nominal Pressure (NP): 9 atm Rated Burst Pressure (RBP): 16 atm (RBP for 4.5mm-14 atm)
Stent Length        Crossing Profile
2.25 mm                 0.82 mm
2.50mm                  0.85 mm
2.75mm                  0.90 mm
3.00 mm                1.00 mm
3.50 mm                1.10 mm
4.00 mm                1.16 mm
4.50 mm                1.20 mm
Balloon Overhang <0.5mm
Dogboning <0.3mm
Guiding catheter Inner Diameter 5 Fr (0.056”) (1.4 mm) Compatible
Catheter Outer Shaft Diameter 2.70 Fr Proximally, 1.95 Fr Distally
Max. Guide wire 0.014 inch